MOUNTING AND STARTUP
1. Select location for display. A location that is out of direct sun-
lightprovidesbestvisibility.
Note:TheTempTrackermaybemountedfromabulkhead
orinstalledinaushpanelbyremovingthemounting
bracket.Panelcutoutdimensionsaregiveninthespeci-
cations.Anoptionalrackmountingpanelisavailablefrom
your YOUNG supplier.
2. Connectcablestoterminals.Refertowiringdiagram.
3. ConnectGNDterminaltosuitableearthground.
4. Insert power supply plug into power jack, plug AC adapter into
wall outlet.
5. TheTempTrackerwilldisplayasoftwareversionnumberfor
severalseconds.Itwillthenbegintodisplaytemperature&
humidity information. The following information is displayed:
• Temperature
• RelativeHumidity%(orDewPoint°)
• Alarmstatus
6. Observethedisplayforafewminutestoverifythatitisoperat-
ing properly. To change settings see the following section.
CHANGING SETTINGS
The Temp Tracker has a setup mode that allows you to easily
change input type, temperature units, alarm settings, and other
functions.
Press and hold ENTER and SELECTkeys(about5seconds).The
displaywillbrieyash“SEtuP”,thenbeginthesetupsequence.
Change settings with the SELECT key. Press the ENTER key to
savesettingsandmovetothenextstep.Abbreviationsintheleft
andrightdisplaywindowsidentifyeachfunctionandtheavailable
selections.
Display Setup Function
LEFT RIGHT
Input Type
InP Prb T/RHprobeinput.
SEr RS-485serialinput(frommaindisplay).
If SEr is selected, setup sequence ends here.
Temperature Units
unit dEG FahrenheitorCelsiusindicatorashes.Select
one.
Relative Humidity/Dew Point
diSP rh DisplayRH%.
dPt Display Dew Point Temperature.
Minimum/Maximum
AUtO no Press SET key for MIN / MAX values.
YES Automatically scroll CURRENT / MIN / MAX
values.
Temperature Alarm
ALr no Temperature alarm not used.
YES Temperature alarm activated.
If YES is selected
SEt 000 T alarm HI setpoint.
SEt 000 T alarm LO setpoint.
Relative Humidity or
Dew Point Alarm
ALr no RH (or DP) alarm not used.
YES RH (or DP) alarm activated.
If YES is selected
SEt 000 RH (or DP) alarm HI setpoint.
SEt 000 RH (or DP) alarm LO setpoint.
If either alarm selected
Alarm delay time
dLY 000 Set alarm delay time in seconds (0-999).
Sound
Snd YES Audiblebeepersoundswithalarm.
no No sound with alarm.
Test Functions
tESt no Skip test functions.
If no is selected, the unit returns to normal operation.
YES Fortroubleshootingonly.YESinitiatesthe
following tests:
tESt Snd PressSELECTtosoundbeeper.
tESt dSP Press SELECT to illuminate all display seg-
ments.
tESt ALr Press SELECT to close alarm relays.
CAL 0.00 PressSELECTtoalternatebetween0.00VDC
and5.00VDCatVoutterminals.Usetocali-
brateexternaldevices(recorders,etc...)
ADDITIONAL INFORMATION
ALARMS
Alarm functions for temperature, humidity (or Dew Point) are ac-
cessedintheSETUPsequence.Eitherorbothalarmsmaybe
selected. Each alarm has a LO and a HI setpoint. When the indi-
catedvaluegoesbelowtheLOorabovetheHIsetpoint,thealarm
indicatorblinks,andtheassociatedrelaycontactcloses.Whena
delay time is set, the indicator reports an alarm condition only after
ithasexistedforonecompletedelayperiod.Alarmactivityceases
when conditions are outside the alarm range for one complete
delay period. For a "latching" alarm effect, use the Temp Tracker
relaycontactstoactivateanexternallatching-typerelay.
BRIGHTNESS
AdjustdisplaybrightnessbyholdingtheBRIGHTkey.
DEW POINT
DewPointisdisplayedintherightdisplaywindowbyselectingthis
feature during SETUP. Dew Point values are calculated from tem-
perature and relative humidity measurements. Dew Point is shown
inthesameunitsasthoseselectedfortemperature(°C,°F).The
"DEWPOINT"labelincludedmaybeplacedover"RELATIVE
HUMIDITY%"onthefrontpanel.